[Design of power and data telemetry system utilizing Class-E amplifier for visual prosthesis].

Authors :
Liu B
Wu K
Wu X
Chai X
Source :
Zhongguo yi liao qi xie za zhi = Chinese journal of medical instrumentation [Zhongguo Yi Liao Qi Xie Za Zhi] 2011 Jul; Vol. 35 (4), pp. 239-42, 245.
Publication Year :
2011

Abstract

A wireless transmission system for visual prosthesis is introduced. Power and data are transmitted through inductive link between only one pair of coils, in order to reduce the size of the device. The transmitter is based on Class-E power amplifier due to its high efficiency, and a modulation circuit is added to control the dc supply of the Class-E amplifier to achieve ASK modulation of data with modulation depth of 25%. In addition, an experimental system is implemented, and test shows that it sufficiently supports a 16-channel micro-stimulator on power delivery and data transfer rate.