[Assessment of uteroplacental circulation in uncomplicated pregnancies using pulsed Doppler ultrasound].

Authors :
Chaoui R
Hoffmann H
Bollmann R
Halle H
Zienert A
Metzner A
Source :
Zentralblatt fur Gynakologie [Zentralbl Gynakol] 1990; Vol. 112 (1), pp. 11-8.
Publication Year :
1990

Abstract

The Doppler sonography enable as a new non-invasive procedure the assessment of the uteroplacental circulation. 209 blood flow measurements (pulsed wave Doppler) were recorded between the 20th and 40th weeks of gestation from the arcuate arteries in 130 women with uncomplicated pregnancy. The S/D Ratio, the Resistance Index (RI) and the Pulsatility Index (PI) were calculated. These Indices show throughout the observation time persisting low values which reflect the optimal uteroplacental circulation in a system with low downstream impedance. We found that the S/D Ratio and the RI are appropriate parameters in the qualitative analysis of the uteroplacental perfusion. On the other side we could demonstrate that the PI depends on the maternal heart rate which explains the high variability of the values. Therefore is the PI not appropriate for the impedance measurements in uteroplacental vessels.