Evaluation of pulsed radiofrequency glow discharge time-of-flight mass spectrometry for precious metal determination in lead fire assay buttons.

Authors :
Compernolle S
Pisonero J
Bordel N
Wambeke D
De Raedt I
Kimpe K
Sanz-Medel A
Vanhaecke F
Source :
Analytica chimica acta [Anal Chim Acta] 2011 Sep 09; Vol. 701 (2), pp. 129-33. Date of Electronic Publication: 2011 Jun 23.
Publication Year :
2011

Abstract

In this paper, an exploration of the capabilities and limitations of pulsed radiofrequency glow discharge time-of-flight mass spectrometry (GD-TOFMS) for the determination of the precious metals Ag, Au, Pd, Pt and Rh in lead buttons obtained by Pb fire assay is reported on. Since the matrix consists almost entirely of lead (>99%), the occurrence of doubly charged Pb (Pb(2+)) ions can hinder accurate determination of Rh. This problem was counteracted by relying on the time-resolved formation of different ion types over the pulse period of the glow discharge, which allows discrimination against the Pb(2+) ions. The formation of ArCu(+) ions as a result of the use of a copper anode was assessed to pose no threat to the accuracy of the results obtained for the set of samples analyzed as its contribution to the total signal at m/z=103 could be adequately corrected for. The method developed was evaluated in terms of accuracy and precision using a set of Pb button standards with analyte concentrations between 5 and 100 μg g(-1). For the purpose of validation, a 10 μg g(-1) standard was considered as a sample. Overall, an acceptable accuracy was obtained with a bias of <5% between the experimental results and the corresponding reference values, except for Au, for which a larger deviation occurred. Precision values (repeatability) of typically <5% relative standard deviation (RSD) (for N=3) were obtained and the limits of detection (LODs) vary from ~0.020 μg g(-1) for Ag to ~0.080 μg g(-1) for Pt.<br /> (Copyright © 2011 Elsevier B.V. All rights reserved.)

Details

Language :
English
ISSN :
1873-4324
Volume :
701
Issue :
2
Database :
MEDLINE
Journal :
Analytica chimica acta
Publication Type :
Academic Journal
Accession number :
21801878
Full Text :
https://doi.org/10.1016/j.aca.2011.05.043