Back to Search Start Over

Determination of serum cholinesterase activity and dibucaine numbers by an amperometric choline sensor.

Authors :
Palleschi G
Lavagnini MG
Moscone D
Pilloton R
D'Ottavio D
Evangelisti ME
Source :
Biosensors & bioelectronics [Biosens Bioelectron] 1990; Vol. 5 (1), pp. 27-35.
Publication Year :
1990

Abstract

Serum cholinesterase activity and the dibucaine numbers have been determined by using a hydrogen peroxide electrode and the enzyme choline oxidase immobilized on a nylon net. The analysis procedure is extremely simple and very fast allowing 30 cholinesterase determinations per hour. Both cholinesterase activity and dibucaine number measurements could be performed in 5 min and by using serum samples of only 10 microliters. When used in sera the probe showed no interference from electroactive compounds present in blood, and also showed good stability and reproducibility. These features make this sensor appropriate for continuous extracorporeal circuit blood monitoring of succinylcholine during surgery.

Details

Language :
English
ISSN :
0956-5663
Volume :
5
Issue :
1
Database :
MEDLINE
Journal :
Biosensors & bioelectronics
Publication Type :
Academic Journal
Accession number :
2178637
Full Text :
https://doi.org/10.1016/0956-5663(90)80024-8