Iterative reconstruction reduces abdominal CT dose.
- Source :
-
European journal of radiology [Eur J Radiol] 2012 Jul; Vol. 81 (7), pp. 1483-7. Date of Electronic Publication: 2011 May 02. - Publication Year :
- 2012
-
Abstract
- Objective: In medical imaging, lowering radiation dose from computed tomography scanning, without reducing diagnostic performance is a desired achievement. Iterative image reconstruction may be one tool to achieve dose reduction. This study reports the diagnostic performance using a blending of 50% statistical iterative reconstruction (ASIR) and filtered back projection reconstruction (FBP) compared to standard FBP image reconstruction at different dose levels for liver phantom examinations.<br />Methods: An anthropomorphic liver phantom was scanned at 250, 185, 155, 140, 120 and 100 mAs, on a 64-slice GE Lightspeed VCT scanner. All scans were reconstructed with ASIR and FBP. Four readers evaluated independently on a 5-point scale 21 images, each containing 32 test sectors. In total 672 areas were assessed. ROC analysis was used to evaluate the differences.<br />Results: There was a difference in AUC between the 250 mAs FBP images and the 120 and 100 mAs FBP images. ASIR reconstruction gave a significantly higher diagnostic performance compared to standard reconstruction at 100 mAs.<br />Conclusion: A blending of 50-90% ASIR and FBP may improve image quality of low dose CT examinations of the liver, and thus give a potential for reducing radiation dose.<br /> (Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1872-7727
- Volume :
- 81
- Issue :
- 7
- Database :
- MEDLINE
- Journal :
- European journal of radiology
- Publication Type :
- Academic Journal
- Accession number :
- 21543174
- Full Text :
- https://doi.org/10.1016/j.ejrad.2011.04.021