Back to Search Start Over

Quantitative analysis of slurry sample by laser-induced breakdown spectroscopy.

Authors :
Ayyalasomayajula KK
Dikshit V
Yueh FY
Singh JP
Smith LT
Source :
Analytical and bioanalytical chemistry [Anal Bioanal Chem] 2011 Jul; Vol. 400 (10), pp. 3315-22. Date of Electronic Publication: 2011 Mar 20.
Publication Year :
2011

Abstract

Laser-induced breakdown spectroscopy (LIBS) has been employed for the analysis of slurry samples. Quantitative analysis of slurry samples is crucial and challenging. The problems associated with slurry samples include splashing, surface turbulence, and the difficulties of obtaining reproducible samples due to sedimentation. The LIBS analysis has achieved limited success due to inherent disadvantages when applied to slurry samples. In order to achieve improved measurement precision and accuracy, a spin-on-glass sampling method was evaluated. Five elements (Al, Ca, Fe, Ni, and Si) were examined in five slurry simulants containing varying amounts of each ion. Three calibration models were developed by using univariate calibration, multiple linear regression, and partial least square regression. LIBS analysis results obtained from the partial least square regression model were determined to be the best fit to results obtained from inductively coupled plasma optical emission spectroscopy analysis.

Details

Language :
English
ISSN :
1618-2650
Volume :
400
Issue :
10
Database :
MEDLINE
Journal :
Analytical and bioanalytical chemistry
Publication Type :
Academic Journal
Accession number :
21424178
Full Text :
https://doi.org/10.1007/s00216-011-4852-3