Back to Search
Start Over
Implementation of alanine/EPR as transfer dosimetry system in a radiotherapy audit programme in Belgium.
- Source :
-
Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology [Radiother Oncol] 2011 Apr; Vol. 99 (1), pp. 94-6. Date of Electronic Publication: 2011 Mar 11. - Publication Year :
- 2011
-
Abstract
- A measurement procedure based on alanine/electron paramagnetic resonance (EPR) dosimetry was implemented successfully providing simple, stable, and accurate dose-to-water (D(w)) measurements. The correspondence between alanine and ionization chamber measurements in reference conditions was excellent. Alanine/EMR dosimetry might be a valuable alternative to thermoluminescent (TLD) and ionization chamber based measuring procedures in radiotherapy audits.<br /> (Copyright © 2011 Elsevier Ireland Ltd. All rights reserved.)
Details
- Language :
- English
- ISSN :
- 1879-0887
- Volume :
- 99
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
- Publication Type :
- Academic Journal
- Accession number :
- 21397968
- Full Text :
- https://doi.org/10.1016/j.radonc.2011.01.026