Interference of gadolinium on the measurement of selenium in human serum by inductively coupled plasma-quadrupole mass spectrometry.
- Source :
-
Annals of clinical biochemistry [Ann Clin Biochem] 2011 Mar; Vol. 48 (Pt 2), pp. 176-7. Date of Electronic Publication: 2011 Feb 25. - Publication Year :
- 2011
-
Abstract
- Selenium is an important clinical biomarker of nutritional status; however, the occurrence of gadolinium in a patient's serum as a result of the contrast agents used during magnetic resonance imaging investigations, results in a significant positive bias in its measurement by inductively coupled plasma-quadrupole mass spectrometry.
Details
- Language :
- English
- ISSN :
- 1758-1001
- Volume :
- 48
- Issue :
- Pt 2
- Database :
- MEDLINE
- Journal :
- Annals of clinical biochemistry
- Publication Type :
- Academic Journal
- Accession number :
- 21355013
- Full Text :
- https://doi.org/10.1258/acb.2010.010030