Predictors of response to a nasal expiratory resistor device and its potential mechanisms of action for treatment of obstructive sleep apnea.
- Source :
-
Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine [J Clin Sleep Med] 2011 Feb 15; Vol. 7 (1), pp. 13-22. - Publication Year :
- 2011
-
Abstract
- Study Objective: A one-way nasal resistor has recently been shown to reduce sleep disordered breathing (SDB) in a subset of patients with Obstructive Sleep Apnea Hypopnea Syndrome (OSAHS). The purpose of this study was to examine characteristics predictive of therapeutic response to the device and provide pilot data as to its potential mechanisms of action. PATIENTS, INTERVENTIONS, AND MEASUREMENTS: 20 subjects (15M/5F, age 54 ± 12 years, BMI 33.5 ± 5.6 kg/m²) with OSAHS underwent 3 nocturnal polysomnograms (NPSG) including diagnostic, therapeutic (with a Provent® nasal valve device), and CPAP. Additional measurements included intranasal pressures and PCO₂, closing pressures (Pcrit), and awake lung volumes in different body positions.<br />Results: In 19/20 patients who slept with the device, RDI was significantly reduced with the nasal valve device compared to the diagnostic NPSG (27 ± 29/h vs 49 ± 28/h), with 50% of patients having an acceptable therapeutic response. Among demographic, lung volume, or diagnostic NPSG measures or markers of collapsibility, no significant predictors of therapeutic response were found. There was a suggestion that patients with position-dependent SDB (supine RDI > lateral RDI) were more likely to have an acceptable therapeutic response to the device. Successful elimination of SDB was associated with generation and maintenance of an elevated end expiratory pressure. No single definitive mechanism of action was elucidated.<br />Conclusions: The present study shows that the nasal valve device can alter SDB across the full spectrum of SDB severity. There was a suggestion that subjects with positional or milder SDB in the lateral position were those most likely to respond.
- Subjects :
- Adult
Continuous Positive Airway Pressure
Equipment Safety
Exhalation
Female
Humans
Male
Middle Aged
Pilot Projects
Polysomnography methods
Posture physiology
Predictive Value of Tests
Prosthesis Design
Respiratory Function Tests
Sampling Studies
Sleep Apnea, Obstructive diagnosis
Tidal Volume
Treatment Outcome
Airway Resistance physiology
Nose
Prostheses and Implants
Sleep Apnea, Obstructive physiopathology
Sleep Apnea, Obstructive therapy
Subjects
Details
- Language :
- English
- ISSN :
- 1550-9397
- Volume :
- 7
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- Journal of clinical sleep medicine : JCSM : official publication of the American Academy of Sleep Medicine
- Publication Type :
- Academic Journal
- Accession number :
- 21344051