Back to Search Start Over

Electromagnetic interference with electrocardiogram recording of exercise test equipment.

Authors :
Aliyev F
Türkoğlu C
Celiker C
Uzunhasan I
Source :
Turk Kardiyoloji Dernegi arsivi : Turk Kardiyoloji Derneginin yayin organidir [Turk Kardiyol Dern Ars] 2010 Jul; Vol. 38 (5), pp. 352-4.
Publication Year :
2010

Abstract

Herein, we report a case of pseudosinus tachycardia resulting from an electromagnetic interference between a mobile phone and treadmill device. Electromagnetic interference from a charging mobile phone connected to the same socket with the exercise device turned the recording of a patient to that of pseudosinus tachycardia at approximately twice the rate of actual basal heart rate. Removal of the mobile phone from the socket resulted in normalization of the electrocardiogram.