The anomalous currents in the front foils of the JET lost alpha diagnostic KA-2.

Authors :
Cecil FE
Kiptily V
Salmi A
Horton A
Fullard K
Murari A
Darrow D
Hill K
Source :
The Review of scientific instruments [Rev Sci Instrum] 2010 Oct; Vol. 81 (10), pp. 10D326.
Publication Year :
2010

Abstract

We have examined the observed currents in the front foils of the JET Faraday cup lost alpha particle diagnostic KA-2. In particular, we have sought to understand the currents during Ohmic plasmas for which the ion flux at the detectors was initially assumed to be negligible. We have considered two sources of this current: plasma ions (both deuterium and impurity) in the vicinity of the detector (including charge exchange neutrals) and photoemission from scattered UV radiation. Based upon modeling and empirical observation, the latter source appears most likely and, moreover, seems to be applicable to the currents in the front foil during ELMy H-mode plasmas. A very thin gold or nickel foil attached to the present detector aperture is proposed as a solution to this problem, and realistic calculations of expected fluxes of lost energetic neutral beam ions during TF ripple experiments are presented as justification of this proposed solution.

Details

Language :
English
ISSN :
1089-7623
Volume :
81
Issue :
10
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
21058456
Full Text :
https://doi.org/10.1063/1.3502040