Back to Search Start Over

Gamma bang time analysis at OMEGA.

Authors :
McEvoy AM
Herrmann HW
Horsfield CJ
Young CS
Miller EK
Mack JM
Kim Y
Stoeffl W
Rubery M
Evans S
Sedillo T
Ali ZA
Source :
The Review of scientific instruments [Rev Sci Instrum] 2010 Oct; Vol. 81 (10), pp. 10D322.
Publication Year :
2010

Abstract

Absolute bang time measurements with the gas Cherenkov detector (GCD) and gamma reaction history (GRH) diagnostic have been performed to high precision at the OMEGA laser facility at the University of Rochester with bang time values for the two diagnostics agreeing to within 5 ps on average. X-ray timing measurements of laser-target coupling were used to calibrate a facility-generated laser timing fiducial with rms spreads in the measured coupling times of 9 ps for both GCD and GRH. Increased fusion yields at the National Ignition Facility (NIF) will allow for improved measurement precision with the GRH easily exceeding NIF system design requirements.

Details

Language :
English
ISSN :
1089-7623
Volume :
81
Issue :
10
Database :
MEDLINE
Journal :
The Review of scientific instruments
Publication Type :
Academic Journal
Accession number :
21033846
Full Text :
https://doi.org/10.1063/1.3485083