Abstract

Multiple reflections between two surfaces of a window introduce a fixed pattern error in optical measurements. One way to remove these spurious reflections is to use a reasonably large wedge so that the interference fringes formed by the two surfaces are too dense for the detector to resolve. However, this method does not work if the wedge angle is small, e.g., several arcseconds. By tilting both the window and the return mirror properly, it is possible to remove the effect of multiple reflections of a window. Theory and experimental results are presented.