An 8×8 Row-Column Summing Readout Electronics for Preclinical Positron Emission Tomography Scanners.

Authors :
Shih YC
Sun FW
Macdonald LR
Otis BP
Miyaoka RS
McDougald W
Lewellen TK
Source :
IEEE Nuclear Science Symposium conference record. Nuclear Science Symposium [IEEE Nucl Sci Symp Conf Rec (1997)] 2009 Oct 24; Vol. 2009, pp. 2376-2380.
Publication Year :
2009

Abstract

This work presents a row/column summing readout electronics for an 8×8 silicon photomultiplier array. The summation circuit greatly reduces the number of electronic channels, which is desirable for pursuing higher resolution positron emission tomography scanners. By using a degenerated common source topology in the summation circuit, more fan-in is possible and therefore a greater reduction in the number of electronic channels can be achieved. The timing signal is retrieved from a common anode, which allows the use of a single fast-sampling analog to digital converter (ADC) for the timing channel and slower, lower power ADCs for the 64 spatial channels. Preliminary results of one row summation of the 8×8 readout electronics exhibited FWHM energy resolution of 17.8% and 18.3% with and without multiplexing, respectively. The measured timing resolution is 2.9ns FWHM.