Miniaturized video-rate epi-third-harmonic-generation fiber-microscope.
- Source :
-
Optics express [Opt Express] 2010 Aug 02; Vol. 18 (16), pp. 17382-91. - Publication Year :
- 2010
-
Abstract
- With a micro-electro-mechanical system (MEMS) mirror, we successfully developed a miniaturized epi-third-harmonic-generation (epi-THG) fiber-microscope with a video frame rate (31 Hz), which was designed for in vivo optical biopsy of human skin. With a large-mode-area (LMA) photonic crystal fiber (PCF) and a regular microscopic objective, the nonlinear distortion of the ultrafast pulses delivery could be much reduced while still achieving a 0.4 microm lateral resolution for epi-THG signals. In vivo real time virtual biopsy of the Asian skin with a video rate (31 Hz) and a sub-micron resolution was obtained. The result indicates that this miniaturized system was compact enough for the least invasive hand-held clinical use.