Accuracy and reproducibility of a novel semi-automatic segmentation technique for MR volumetry of the pituitary gland.

Authors :
Renz DM
Hahn HK
Schmidt P
Rexilius J
Lentschig M
Pfeil A
Sauner D
Fitzek C
Mentzel HJ
Kaiser WA
Reichenbach JR
Böttcher J
Source :
Neuroradiology [Neuroradiology] 2011 Apr; Vol. 53 (4), pp. 233-44. Date of Electronic Publication: 2010 Jun 19.
Publication Year :
2011

Abstract

Introduction: Although several reports about volumetric determination of the pituitary gland exist, volumetries have been solely performed by indirect measurements or manual tracing on the gland's boundaries. The purpose of this study was to evaluate the accuracy and reproducibility of a novel semi-automatic MR-based segmentation technique.<br />Methods: In an initial technical investigation, T1-weighted 3D native magnetised prepared rapid gradient echo sequences (1.5 T) with 1 mm isotropic voxel size achieved high reliability and were utilised in different in vitro and in vivo studies. The computer-assisted segmentation technique was based on an interactive watershed transform after resampling and gradient computation. Volumetry was performed by three observers with different software and neuroradiologic experiences, evaluating phantoms of known volume (0.3, 0.9 and 1.62 ml) and healthy subjects (26 to 38 years; overall 135 volumetries).<br />Results: High accuracy of the volumetry was shown by phantom analysis; measurement errors were <4% with a mean error of 2.2%. In vitro, reproducibility was also promising with intra-observer variability of 0.7% for observer 1 and 0.3% for observers 2 and 3; mean inter-observer variability was in vitro 1.2%. In vivo, scan-rescan, intra-observer and inter-observer variability showed mean values of 3.2%, 1.8% and 3.3%, respectively. Unifactorial analysis of variance demonstrated no significant differences between pituitary volumes for various MR scans or software calculations in the healthy study groups (p > 0.05).<br />Conclusion: The analysed semi-automatic MR volumetry of the pituitary gland is a valid, reliable and fast technique. Possible clinical applications are hyperplasia or atrophy of the gland in pathological circumstances either by a single assessment or by monitoring in follow-up studies.

Details

Language :
English
ISSN :
1432-1920
Volume :
53
Issue :
4
Database :
MEDLINE
Journal :
Neuroradiology
Publication Type :
Academic Journal
Accession number :
20563571
Full Text :
https://doi.org/10.1007/s00234-010-0727-0