Back to Search Start Over

Characteristics of a passive RF field probe with fibre-optic link for measurements in liquid hyperthermia phantoms.

Authors :
Schneider CJ
Engelberts N
van Dijk JD
Source :
Physics in medicine and biology [Phys Med Biol] 1991 Apr; Vol. 36 (4), pp. 461-74.
Publication Year :
1991

Abstract

An experimental study is described, leading to the design of a near-field electromagnetic radiation probe. A passive field probe with fibre-optic link has been constructed for use in hyperthermia phantoms, i.e. for measurements in lossy media--saline was used in this study as a simulation of biological tissue. The realization of a closed electric circuit on the probe proved to be a major influence on the sensitivity and dynamic range of the probe. The experiments also independently confirmed the dipole length, frequency variations between 5 and 110 MHz and the temperature of the probe as important parameters.

Details

Language :
English
ISSN :
0031-9155
Volume :
36
Issue :
4
Database :
MEDLINE
Journal :
Physics in medicine and biology
Publication Type :
Academic Journal
Accession number :
2047396
Full Text :
https://doi.org/10.1088/0031-9155/36/4/005