Back to Search Start Over

Transapical transcatheter aortic valve implantation: follow-up to 3 years.

Authors :
Ye J
Cheung A
Lichtenstein SV
Nietlispach F
Albugami S
Masson JB
Thompson CR
Munt B
Moss R
Carere RG
Jamieson WR
Webb JG
Source :
The Journal of thoracic and cardiovascular surgery [J Thorac Cardiovasc Surg] 2010 May; Vol. 139 (5), pp. 1107-13, 1113.e1.
Publication Year :
2010

Abstract

Background: We performed the first human case of successful transapical transcatheter aortic valve implantation on a beating heart in October 2005, and therefore we have the longest follow-up on transapical aortic valve implantation in humans. We now report clinical and echocardiographic outcomes of transapical aortic valve implantation in 71 patients.<br />Methods: Between October 2005 and February 2009, 71 patients (44 female) underwent transcatheter transapical aortic valve implantation with either 23- or 26-mm Edwards Lifesciences transcatheter bioprostheses. All patients with symptomatic aortic stenosis were declined for conventional aortic valve replacement owing to unacceptable operative risks and were not candidates for transfemoral aortic valve implantation because of poor arterial access. Clinical and echocardiographic follow-ups were performed before discharge, at 1 and 6 months, and then yearly. The mean follow-up was 12.9 +/- 11.5 months with a total of 917.3 months of follow-up.<br />Results: Mean age was 80.0 +/- 8.1 years and predicted operative mortality was 34.5% +/- 20.4% by logistic EuroSCORE and 12.1% +/- 7.7% by The Society of Thoracic Surgeons Risk Calculator. Valves were successfully implanted in all patients. Twelve patients died within 30 days (30-day mortality: 16.9% in all patients, 33% in the first 15 patients, and 12.5% in the remainder), and 10 patients died subsequently. Overall survival at 24 and 36 months was 66.3% +/- 6.4% and 58.0% +/- 9.5%, respectively. Among 59 patients who survived at least 30 days, 24- and 36-month survivals were 79.8% +/- 6.4% and 69.8% +/- 10.9%, respectively. Late valve-related complications were rare. New York Heart Association functional class improved significantly from preoperative 3.3 +/- 0.8 to 1.8 +/- 0.8 at 24 months. The aortic valve area and mean gradient remained stable at 24 months (1.6 +/- 0.3 cm(2) and 10.3 +/- 5.9 mm Hg, respectively).<br />Conclusion: Our outcome suggests that transapical transcatheter aortic valve implantation provides sustained clinical and hemodynamic benefits for up to 36 months in selected high-risk patients with symptomatic severe aortic stenosis.<br /> (2010 The American Association for Thoracic Surgery. Published by Mosby, Inc. All rights reserved.)

Details

Language :
English
ISSN :
1097-685X
Volume :
139
Issue :
5
Database :
MEDLINE
Journal :
The Journal of thoracic and cardiovascular surgery
Publication Type :
Academic Journal
Accession number :
20412948
Full Text :