Back to Search Start Over

Late failure of a single-coil transvenous implantable cardioverter-defibrillator lead associated with conductor separation.

Authors :
Richards MW
Warren CE
Anderson MH
Source :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology [Europace] 2010 Aug; Vol. 12 (8), pp. 1191-2. Date of Electronic Publication: 2010 Mar 27.
Publication Year :
2010

Abstract

Two patients with the same model of single-coil active fix implantable defibrillator lead presented with evidence of lead malfunction, in one case with an abrupt rise in pacing impedance and threshold and in the other with evidence of over-sensing. In both cases, the chest radiograph showed separation of conductors in the defibrillation lead from the main body of the lead.

Details

Language :
English
ISSN :
1532-2092
Volume :
12
Issue :
8
Database :
MEDLINE
Journal :
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology
Publication Type :
Academic Journal
Accession number :
20348555
Full Text :
https://doi.org/10.1093/europace/euq082