Back to Search
Start Over
Late failure of a single-coil transvenous implantable cardioverter-defibrillator lead associated with conductor separation.
- Source :
-
Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology [Europace] 2010 Aug; Vol. 12 (8), pp. 1191-2. Date of Electronic Publication: 2010 Mar 27. - Publication Year :
- 2010
-
Abstract
- Two patients with the same model of single-coil active fix implantable defibrillator lead presented with evidence of lead malfunction, in one case with an abrupt rise in pacing impedance and threshold and in the other with evidence of over-sensing. In both cases, the chest radiograph showed separation of conductors in the defibrillation lead from the main body of the lead.