An injection method for measuring the carbon isotope content of soil carbon dioxide and soil respiration with a tunable diode laser absorption spectrometer.

Authors :
Moyes AB
Schauer AJ
Siegwolf RT
Bowling DR
Source :
Rapid communications in mass spectrometry : RCM [Rapid Commun Mass Spectrom] 2010 Apr 15; Vol. 24 (7), pp. 894-900.
Publication Year :
2010

Abstract

We present a novel technique in which the carbon isotope ratio (delta(13)C) of soil CO(2) is measured from small gas samples (<5 mL) injected into a stream of CO(2)-free air flowing into a tunable diode laser absorption spectrometer (TDL). This new method extends the dynamic range of the TDL to measure CO(2) mole fractions ranging from ambient to pure CO(2), reduces the volume of sample required to a few mL, and does not require field deployment of the instrument. The measurement precision of samples stored for up to 60 days was 0.23 per thousand. The new TDL method was applied with a simple gas well sampling technique to obtain and measure gas samples from shallow soil depth increments for CO(2) mole fraction and delta(13)C analysis, and subsequent determination of the delta(13)C of soil-respired CO(2). The method was tested using an artificial soil system containing a controlled CO(2) source and compared with an independent method using the TDL and an open soil chamber. The profile and chamber estimates of delta(13)C of an artificially produced CO(2) flux were consistent and converged to the delta(13)C of the CO(2) source at steady state, indicating the accuracy of both methods under controlled conditions. The new TDL method, in which a small pulse of sample is measured on a carrier gas stream, is analogous for the TDL technique to the development of continuous-flow configurations for isotope ratio mass spectrometry. While the applications presented here are focused on soil CO(2), this new TDL method could be applied in a number of situations requiring measurement of delta(13)C of CO(2) in small gas samples with ambient to high CO(2) mole fractions.<br /> (2010 John Wiley & Sons, Ltd.)

Details

Language :
English
ISSN :
1097-0231
Volume :
24
Issue :
7
Database :
MEDLINE
Journal :
Rapid communications in mass spectrometry : RCM
Publication Type :
Academic Journal
Accession number :
20201033
Full Text :
https://doi.org/10.1002/rcm.4466