Back to Search
Start Over
In vitro biomechanical analysis of a new lumbar low-profile locking screw-plate construct versus a standard top-loading cantilevered pedicle screw-rod construct: technical report.
- Source :
-
Neurosurgery [Neurosurgery] 2010 Feb; Vol. 66 (2), pp. E404-6; discussion E406. - Publication Year :
- 2010
-
Abstract
- Objective: A standard top-loading lumbar pedicle screw-rod system is compared with a pedicle screw-plate system with smaller-diameter screws, more medial entry, and lower profile to assess the relative stability, strength, and resistance to fatigue of the 2 systems.<br />Methods: Seven human cadaveric specimens were studied with each surgical construct. Nondestructive, nonconstraining pure moments were applied to specimens to induce flexion, extension, lateral bending, and axial rotation while recording L5-S1 motion optoelectronically. After initial tests, specimens were fatigued for 10,000 cycles and retested to assess early postoperative loosening. Specimens were then loaded to failure in hyperextension.<br />Results: The standard screw-rod construct reduced range of motion to a mean of 20% of normal, whereas the screw-plate construct reduced range of motion to 13% of normal. Differences between systems were not significant in any loading mode (P > 0.06). The 14% loosening of the screw-rod system with fatigue was not significantly different from the 10% loosening observed with the screw-plate system (P > 0.15). Mean failure loads of 30 Nm for screw-rod and 37 Nm for screw-plate were also not significantly different (P = 0.38).<br />Conclusion: Posterior fixation at L5-S1 using the low-profile screw-plate system offers stability, resistance to fatigue, and resistance to failure equivalent to fixation using a standard cantilevered pedicle screw-rod system.
Details
- Language :
- English
- ISSN :
- 1524-4040
- Volume :
- 66
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Neurosurgery
- Publication Type :
- Academic Journal
- Accession number :
- 20087110
- Full Text :
- https://doi.org/10.1227/01.NEU.0000363701.76835.BF