Back to Search Start Over

Introducer needles of peripheral intravenous catheters: assessment of magnetic field interactions with 1.5T and 3T MR systems.

Authors :
Sakai M
Aoki S
Watanabe Y
Tanabe D
Taga T
Inoue Y
Ohtomo K
Nakamura H
Source :
Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine [Magn Reson Med Sci] 2009; Vol. 8 (4), pp. 181-5.
Publication Year :
2009

Abstract

We developed a peripheral intravenous catheter introducer that can be used safely in the magnetic resonance (MR) environment, including that at 3.0-tesla. We evaluated introducers with stainless steel (SUS 316L) and nickel-chromium-based (inconel 600) needles as well as a 20-gauge peripheral intravenous catheter introducer with SUS 304 needle for MR safety. From an MR safety standpoint, the SUS 304 should not be selected, and though inconel 600 is the preferred material, the SUS 316L introducer may be more practical with some modifications.

Details

Language :
English
ISSN :
1880-2206
Volume :
8
Issue :
4
Database :
MEDLINE
Journal :
Magnetic resonance in medical sciences : MRMS : an official journal of Japan Society of Magnetic Resonance in Medicine
Publication Type :
Academic Journal
Accession number :
20035127
Full Text :
https://doi.org/10.2463/mrms.8.181