Back to Search Start Over

Improving the trade-off between simulation time and accuracy in efficiency calibrations with the code DETEFF.

Authors :
Díaz NC
Vargas MJ
Source :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine [Appl Radiat Isot] 2010 Jul-Aug; Vol. 68 (7-8), pp. 1413-7. Date of Electronic Publication: 2009 Nov 17.
Publication Year :
2010

Abstract

Quick and relatively simple procedures were incorporated into the Monte Carlo code DETEFF in order to consider the escape of Bremsstrahlung radiation and secondary electrons. The relative bias in efficiency values was thus reduced for photon energies between 1500 and 2000 keV, without any noticeable increment of the simulation time. A relatively simple method was also included to consider the rounding of detector edges. The validation studies showed relative deviations of about 1% in the energy range 10-2000 keV.<br /> (Copyright 2009 Elsevier Ltd. All rights reserved.)

Details

Language :
English
ISSN :
1872-9800
Volume :
68
Issue :
7-8
Database :
MEDLINE
Journal :
Applied radiation and isotopes : including data, instrumentation and methods for use in agriculture, industry and medicine