Back to Search Start Over

An integrated circuit for wireless ambulatory arrhythmia monitoring systems.

Authors :
Kim H
Yazicioglu RF
Torfs T
Merken P
Van Hoof C
Yoo HJ
Source :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference [Annu Int Conf IEEE Eng Med Biol Soc] 2009; Vol. 2009, pp. 5409-12.
Publication Year :
2009

Abstract

An ECG signal processor (ESP) is proposed for the low energy wireless ambulatory arrhythmia monitoring system. The ECG processor mainly performs filtering, compression, classification and encryption. The data compression flow consisting of skeleton and modified Huffman coding is the essential function to reduce the transmission energy consumption and the memory capacity, which are the most energy consuming part. The classification flow performs the arrhythmia analysis to alert the abnormality. The proposed ESP IC is implemented in 0.18-microm CMOS process and integrated into the wireless arrhythmia monitoring sensor platform. By integration of the ESP, the total system energy reduction is evaluated by 95.6%.

Details

Language :
English
ISSN :
2375-7477
Volume :
2009
Database :
MEDLINE
Journal :
Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Publication Type :
Academic Journal
Accession number :
19963908
Full Text :
https://doi.org/10.1109/IEMBS.2009.5332815