Secondary charged particle equilibrium in 137Cs and 60Co reference radiation fields.

Authors :
Behrens R
Kowatari M
Hupe O
Source :
Radiation protection dosimetry [Radiat Prot Dosimetry] 2009 Sep; Vol. 136 (3), pp. 168-75. Date of Electronic Publication: 2009 Sep 15.
Publication Year :
2009

Abstract

During the calibration or irradiation of dosemeters, typical irradiation geometries (collimated beams) and source-to-detector distances (1-5 m) lead to the fact that for photon energies above a few hundred keV, the secondary charged particle equilibrium is usually not ensured. The reason is that the effective beam radius at the detector position is smaller than the range of the secondary electrons produced in air whose maximum particle energy is as large as the maximum photon energy. Therefore, the International Organization for Standardization (ISO) recommends putting a build-up plate (BUP) made of polymethyl methacrylate in front of the dosemeter to be calibrated in ISO 4037-3. In this paper, the effect of the thickness of the BUP and its distance from the dosemeter at different source-to-dosemeter distances were investigated by means of measurement and calculation. It turned out that the geometrical arrangement of the source, dosemeter and BUP recommended by ISO mostly does not ensure secondary charged particle equilibrium. The consequence is to always place the BUP directly in front of the dosemeter to be calibrated or irradiated.

Details

Language :
English
ISSN :
1742-3406
Volume :
136
Issue :
3
Database :
MEDLINE
Journal :
Radiation protection dosimetry
Publication Type :
Academic Journal
Accession number :
19755433
Full Text :
https://doi.org/10.1093/rpd/ncp173