Compound analysis of gallstones using dual energy computed tomography--results in a phantom model.
- Source :
-
European journal of radiology [Eur J Radiol] 2010 Jul; Vol. 75 (1), pp. e74-80. Date of Electronic Publication: 2009 Aug 29. - Publication Year :
- 2010
-
Abstract
- Purpose: The potential of dual energy computed tomography (DECT) for the analysis of gallstone compounds was investigated. The main goal was to find parameters, that can reliably define high percentage (>70%) cholesterol stones without calcium components.<br />Materials and Methods: 35 gallstones were analyzed with DECT using a phantom model. Stone samples were put into specimen containers filled with formalin. Containers were put into a water-filled cylindrical acrylic glass phantom. DECT scans were performed using a tube voltage/current of 140 kV/83 mAs (tube A) and 80 kV/340 mAs (tube B). ROI-measurements to determine CT attenuation of each sector of the stones that had different appearance on the CT images were performed. Finally, semi-quantitative infrared spectroscopy (FTIR) of these sectors was performed for chemical analysis.<br />Results: ROI-measurements were performed in 45 different sectors in 35 gallstones. Sectors containing >70% of cholesterol and no calcium component (n=20) on FTIR could be identified with 95% sensitivity and 100% specificity on DECT. These sectors showed typical attenuation of -8+/-4 HU at 80 kV and +22+/-3 HU at 140 kV. Even the presence of a small calcium component (<10%) hindered the reliable identification of cholesterol components as such.<br />Conclusion: Dual energy CT allows for reliable identification of gallstones containing a high percentage of cholesterol and no calcium component in this pre-clinical phantom model. Results from in vivo or anthropomorphic phantom trials will have to confirm these results. This may enable the identification of patients eligible for non-surgical treatment options in the future.<br /> (Copyright 2009 Elsevier Ireland Ltd. All rights reserved.)
- Subjects :
- Humans
Phantoms, Imaging
Radiography, Dual-Energy Scanned Projection instrumentation
Reproducibility of Results
Sensitivity and Specificity
Tomography, X-Ray Computed instrumentation
Gallstones diagnostic imaging
Radiographic Image Enhancement methods
Radiographic Image Interpretation, Computer-Assisted methods
Radiography, Dual-Energy Scanned Projection methods
Tomography, X-Ray Computed methods
Subjects
Details
- Language :
- English
- ISSN :
- 1872-7727
- Volume :
- 75
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- European journal of radiology
- Publication Type :
- Academic Journal
- Accession number :
- 19717259
- Full Text :
- https://doi.org/10.1016/j.ejrad.2009.08.004