Back to Search
Start Over
Comparison of a new reusable gynecologic laparoscopic electric morcellator with a disposable morcellator: a preliminary trial.
- Source :
-
Journal of minimally invasive gynecology [J Minim Invasive Gynecol] 2009 Sep-Oct; Vol. 16 (5), pp. 595-8. Date of Electronic Publication: 2009 Jul 12. - Publication Year :
- 2009
-
Abstract
- We compared the efficiency and safety of a newly developed reusable morcellator (Olympus VarioCarve) with that of a disposable morcellator (GyneCare Morcellex) for gynecologic laparoscopic surgery. In group 1 (n=15 patients; 11 myomectomies and 4 supracervical hysterectomies), the laparoscopic disposable morcellator was used, and in group 2 (n=14; 11 myomectomies and 3 supracervical hysterectomies), the reusable morcellator was used. There were no statistical differences in total tissue weight. Morcellating time was significantly shorter in group 2, and the rate of morcellation was significantly greater in group 2. Significantly fewer and longer pieces of tissue were removed with the reusable morcellator. No iatrogenic organ damage was observed.
Details
- Language :
- English
- ISSN :
- 1553-4650
- Volume :
- 16
- Issue :
- 5
- Database :
- MEDLINE
- Journal :
- Journal of minimally invasive gynecology
- Publication Type :
- Academic Journal
- Accession number :
- 19596217
- Full Text :
- https://doi.org/10.1016/j.jmig.2009.05.010