MR-compatible polyetheretherketone-based guide wire assisting MR-guided stenting of iliac and supraaortic arteries in swine: feasibility study.
- Source :
-
Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy [Minim Invasive Ther Allied Technol] 2009; Vol. 18 (3), pp. 181-8. - Publication Year :
- 2009
-
Abstract
- The purpose of this study was to demonstrate first magnetic resonance (MR)-guided stenting of iliac and supraaortic arteries using a polyetheretherketone-based (PEEK) MR-compatible guide wire. In vitro and animal experiments were performed in a short magnet wide-bore scanner (1.5 Tesla, Espree, Siemens Healthcare, Erlangen, Germany). For all experiments, a 0.035'' MR-compatible guide wire prototoype was used. This wire had a compound core of PEEK with reinforcing fibres, a soft and atraumatic tip and a hydrophilic coating. For its passive visualization, paramagnetic markings were attached. All experiments were performed through a vascular introducer sheath under MR-guidance. In vitro repetitive selective over the wire catheterizations of either the right carotid artery and the left subclavian artery were performed. In vivo, selective catheterization and over-the-wire stenting of the brachiocephalic trunk and the left subclavian artery were performed. The common iliac arteries were catheterized retrogradely (left) and cross-over (right). Angioplasty and stenting were performed over-the-wire. All procedures were successful. Visibility of the PEEK-based guide-wire was rated good in vitro and acceptable in vivo. Guide wire pushability and endovascular device support were good. The PEEK-based MR-compatible guide wire is well visible and usable under MR-guidance. It supports over-the-wire treatment of iliac arteries and supraaortic arteries.
- Subjects :
- Angioplasty methods
Animals
Aorta surgery
Benzophenones
Biocompatible Materials
Brachiocephalic Trunk surgery
Cardiac Catheterization instrumentation
Cardiac Catheterization methods
Equipment Design
Female
Iliac Artery surgery
Polymers
Subclavian Artery surgery
Swine
Ketones
Magnetic Resonance Imaging, Interventional methods
Polyethylene Glycols
Stents
Subjects
Details
- Language :
- English
- ISSN :
- 1365-2931
- Volume :
- 18
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Minimally invasive therapy & allied technologies : MITAT : official journal of the Society for Minimally Invasive Therapy
- Publication Type :
- Academic Journal
- Accession number :
- 19431070
- Full Text :
- https://doi.org/10.1080/13645700902921971