Back to Search
Start Over
Performance evaluation of a miniature ion mobility spectrometer drift cell for application in hand-held explosives detection ion mobility spectrometers.
- Source :
-
Analytical and bioanalytical chemistry [Anal Bioanal Chem] 2009 Sep; Vol. 395 (2), pp. 411-9. Date of Electronic Publication: 2009 May 08. - Publication Year :
- 2009
-
Abstract
- The implementation of hand-held ion mobility spectrometers (IMS) requires the development and evaluation of miniature drift cells providing high sensitivity while maintaining reasonable resolution. This manuscript describes the construction of a miniature IMS designed for such an application and its characterization by evaluation of the detection limits and resolution of the system with seven explosive compounds including trinitrotoluene (TNT), cyclotrimethylenetrinitramine (RDX), pentaerythritol tetranitrate (PETN), 2,4,6-trinitrophenyl-N-methylnitramine (Tetryl), nitroglycerin (NG), 2,4-dinitrotoluene (2,4 DNT), and 2,6-dinitrotoluene (2,6 DNT).
Details
- Language :
- English
- ISSN :
- 1618-2650
- Volume :
- 395
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Analytical and bioanalytical chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 19424683
- Full Text :
- https://doi.org/10.1007/s00216-009-2818-5