Enhancing dynamic scanning force microscopy in air: as close as possible.
- Source :
-
Nanotechnology [Nanotechnology] 2009 Feb 25; Vol. 20 (8), pp. 085707. Date of Electronic Publication: 2009 Feb 03. - Publication Year :
- 2009
-
Abstract
- Frequency modulation dynamic scanning force microscopy has been implemented in ambient conditions using low oscillation amplitudes (<1 nm) to simultaneously record not only topographic but also additional channels of information, in particular contact potential images. The performance of this mode as compared to the conventional amplitude modulation mode is analyzed in detail using a biological molecule, turning yellow mosaic virus RNA, as the model sample. On the basis of scanning force microscopy imaging as well as spectroscopy experiments, we find that for such very small samples the frequency modulation mode is superior since it can be operated with smaller tip-sample interaction, smaller effective tip-sample distance and lower forces. Combined with Kelvin probe microscopy it results not only in considerably higher electrostatic resolution, but also in correct quantitative values for the contact potential as compared to traditional amplitude modulation scanning force microscopy.
Details
- Language :
- English
- ISSN :
- 1361-6528
- Volume :
- 20
- Issue :
- 8
- Database :
- MEDLINE
- Journal :
- Nanotechnology
- Publication Type :
- Academic Journal
- Accession number :
- 19417468
- Full Text :
- https://doi.org/10.1088/0957-4484/20/8/085707