Back to Search Start Over

A new self-expanding aortic stent valve with annular fixation: in vitro haemodynamic assessment.

Authors :
Vergnat M
Henaine R
Kalejs M
Bommeli S
Ferrari E
Obadia JF
Von Segesser LK
Source :
European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery [Eur J Cardiothorac Surg] 2009 Jun; Vol. 35 (6), pp. 970-5; discussion 975-6. Date of Electronic Publication: 2009 May 05.
Publication Year :
2009

Abstract

Objective: Balloon-expandable stent valves require flow reduction during implantation (rapid pacing). The present study was designed to compare a self-expanding stent valve with annular fixation versus a balloon-expandable stent valve.<br />Methods: Implantation of a new self-expanding stent valve with annular fixation (Symetis, Lausanne, Switzerland) was assessed versus balloon-expandable stent valve, in a modified Dynatek Dalta pulse duplicator (sealed port access to the ventricle for transapical route simulation), interfaced with a computer for digital readout, carrying a 25 mm porcine aortic valve. The cardiovascular simulator was programmed to mimic an elderly woman with aortic stenosis: 120/85 mmHg aortic pressure, 60 strokes/min (66.5 ml), 35% systole (2.8 l/min).<br />Results: A total of 450 cardiac cycles was analysed. Stepwise expansion of the self-expanding stent valve with annular fixation (balloon-expandable stent valve) resulted in systolic ventricular increase from 120 to 121 mmHg (126 to 830+/-76 mmHg)*, and left ventricular outflow obstruction with mean transvalvular gradient of 11+/-1.5 mmHg (366+/-202 mmHg)*, systolic aortic pressure dropped distal to the valve from 121 to 64.5+/-2 mmHg (123 to 55+/-30 mmHg) N.S., and output collapsed to 1.9+/-0.06 l/min (0.71+/-0.37 l/min* (before complete obstruction)). No valve migration occurred in either group. (*=p<0.05).<br />Conclusions: Implantation of this new self-expanding stent valve with annular fixation has little impact on haemodynamics and has the potential for working heart implantation in vivo. Flow reduction (rapid pacing) is not necessary.

Details

Language :
English
ISSN :
1873-734X
Volume :
35
Issue :
6
Database :
MEDLINE
Journal :
European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery
Publication Type :
Academic Journal
Accession number :
19409805
Full Text :
https://doi.org/10.1016/j.ejcts.2009.01.052