Back to Search Start Over

Ionic effect investigation of a potentiometric sensor for urea and surface morphology observation of entrapped urease/polypyrrole matrix.

Authors :
Syu MJ
Chang YS
Source :
Biosensors & bioelectronics [Biosens Bioelectron] 2009 Apr 15; Vol. 24 (8), pp. 2671-7. Date of Electronic Publication: 2009 Feb 02.
Publication Year :
2009

Abstract

Potentio-dynamic polymerization of buffered urease and pyrrole monomer onto carbon papers was conducted to fabricate an immobilized urease electrode for measuring the urea concentration. To use carbon paper as the substrate for the electro-growth of polypyrrole matrix not only created sufficient adhesion of the conducting polymer layer but also provided superior entrapment of urease enzymes. The potentiometric response corresponding to ammonia, the product formed from the urease catalyzed urea reaction, was employed for the urea concentration measurement. Scanning electron microscopic photographs showed that the polypyrrole matrix deposited on the carbon papers appeared to be of a cylindrical nanotube shape. The charge density applied in the polymerization was found to affect the potentiometric response while the potential-scanning rate showed minor influence. The composite electrodes had high sensitivity in urea detection, showing a response linear to the logarithm of the urea concentration in the range of 10(-3) to 10 mM. The detection of urea solution prepared in water and buffer was also compared. Ionic effect on the sensing of urea solution was investigated. By comparing the data reported in literature, the urease/polypyrrole/carbon paper electrode developed in this work showed superior long-term stability and reusability. The detection of urea in serum was also well performed.

Details

Language :
English
ISSN :
1873-4235
Volume :
24
Issue :
8
Database :
MEDLINE
Journal :
Biosensors & bioelectronics
Publication Type :
Academic Journal
Accession number :
19237276
Full Text :
https://doi.org/10.1016/j.bios.2009.01.036