Abstract

Background: Computer- and robot-assisted technologies are capable of improving the accuracy of planar cutting in orthopaedic surgery. This study is a first step toward formulating and validating a new evaluation methodology for planar bone cutting, based on the standards from the International Organization for Standardization.<br />Methods: Our experimental test bed consisted of a purely geometrical model of the cutting process around a simulated bone. Cuts were performed at three levels of surgical assistance: unassisted, computer-assisted and robot-assisted. We measured three parameters of the standard ISO1101:2004: flatness, parallelism and location of the cut plane.<br />Results: The location was the most relevant parameter for assessing cutting errors. The three levels of assistance were easily distinguished using the location parameter.<br />Conclusions: Our ISO methodology employs the location to obtain all information about translational and rotational cutting errors. Location may be used on any osseous structure to compare the performance of existing assistance technologies.

Details

Language :
English
ISSN :
1478-596X
Volume :
5
Issue :
1
Database :
MEDLINE
Journal :
The international journal of medical robotics + computer assisted surgery : MRCAS
Publication Type :
Academic Journal
Accession number :
19172588
Full Text :
https://doi.org/10.1002/rcs.237