Back to Search Start Over

Quantitative analysis of RU38486 (mifepristone) by HPLC triple quadrupole mass spectrometry.

Authors :
Homer NZ
Reynolds RM
Mattsson C
Bailey MA
Walker BR
Andrew R
Source :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences [J Chromatogr B Analyt Technol Biomed Life Sci] 2009 Feb 15; Vol. 877 (5-6), pp. 497-501. Date of Electronic Publication: 2008 Dec 27.
Publication Year :
2009

Abstract

A sensitive liquid chromatography-mass spectrometric method was validated for the quantification of RU38486 (mifepristone) in human and murine plasma. The analyte and internal standard (alfaxolone) were extracted by liquid-liquid extraction with diethyl ether, resolved on a C18 column using gradient elution with methanol and ammonium acetate and detected after positive electrospray ionization (m/z 430-->372; m/z 333-->297, respectively). Quantification was linear over the range 0.5-500ng (r(2)>0.997), precise and accurate (intra-assay RSD< or =7.2%, RME< or =8.2%; inter-assay RSD< or =15.7% RME< or =10.2%). The limit of quantification (LOQ) was 50pg injected on column, permitting reproducible analysis of RU38486 in small volumes of plasma.

Details

Language :
English
ISSN :
1570-0232
Volume :
877
Issue :
5-6
Database :
MEDLINE
Journal :
Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
Publication Type :
Academic Journal
Accession number :
19157997
Full Text :
https://doi.org/10.1016/j.jchromb.2008.12.051