Back to Search Start Over

Detection apparatus for multiple heterogeneous chemiluminescence immunoassay configurations.

Authors :
Khalil OS
Zurek TF
Pepe C
Genger K
Huff DG
Coleman C
Hanna C
Hu R
Mackowiack J
Bennett L
Source :
Analytical biochemistry [Anal Biochem] 1991 Jul; Vol. 196 (1), pp. 61-8.
Publication Year :
1991

Abstract

We describe an apparatus for measuring signals emanated from two heterogeneous chemiluminescence immunoassay (CLIA) configurations: antibody-coated polystyrene beads, in reaction tray wells, and microparticles captured by a porous matrix. An optics and fluidics design which allows the use of a common detection head for these two different assay configurations is described. The detection head moves along three Cartesian coordinates to create a localized light-tight compartment around each individual disposable reaction vessel. Reproducibility of the light seal, trigger solution delivery, and mixing is achieved for acridinium-labeled CLIA. The coated polystyrene beads configuration is tested using beta HCG, CEA, and TSH assays. The microparticle-capture configuration is tested using beta HCG and HBsAg assays. The microparticle capture CLIA has shorter incubation times and the potential for ease of automation.