Back to Search
Start Over
Limited field of view spin echo MR imaging.
- Source :
-
Magnetic resonance imaging [Magn Reson Imaging] 1991; Vol. 9 (3), pp. 389-94. - Publication Year :
- 1991
-
Abstract
- Several groups have reported using a method of limiting the field of view (FOV) where the slices excited by the 90 and 180 degree pulses are perpendicular. However, only one slice can be excited during each repetition time, so multislice imaging is not possible. We present a modification of this method that allows multislice imaging. The slices excited by the 90 degrees and 180 degrees pulses are at a small angle; the field of view is limited and multislice imaging is possible. The modifications also allow the center of the FOV to be offset to any position. We describe the conditions that yield optimal images for the given FOV, slice thickness, and interslice gap. Representative images demonstrating the features of the technique are presented. The technique can be used to reduce the number of phase-encoding steps resulting in reduced imaging time, or it can be used to increase the spatial resolution without increasing the imaging time.
Details
- Language :
- English
- ISSN :
- 0730-725X
- Volume :
- 9
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Magnetic resonance imaging
- Publication Type :
- Academic Journal
- Accession number :
- 1881258
- Full Text :
- https://doi.org/10.1016/0730-725x(91)90427-n