Back to Search Start Over

Biomechanical comparison of single-row arthroscopic rotator cuff repair technique versus transosseous repair technique.

Authors :
Tocci SL
Tashjian RZ
Leventhal E
Spenciner DB
Green A
Fleming BC
Source :
Journal of shoulder and elbow surgery [J Shoulder Elbow Surg] 2008 Sep-Oct; Vol. 17 (5), pp. 808-14. Date of Electronic Publication: 2008 Jul 02.
Publication Year :
2008

Abstract

This study determined the effect of tear size on gap formation of single-row simple-suture arthroscopic rotator cuff repair (ARCR) vs transosseous Mason-Allen suture open RCR (ORCR) in 13 pairs of human cadaveric shoulders. A massive tear was created in 6 pairs and a large tear in 7. Repairs were cyclically tested in low-load and high-load conditions, with no significant difference in gap formation. Under low-load, gapping was greater in massive tears. Under high-load, there was a trend toward increased gap with ARCR for large tears. All repairs of massive tears failed in high-load. Gapping was greater posteriorly in massive tears for both techniques. Gap formation of a modeled RCR depends upon the tear size. ARCR of larger tears may have higher failure rates than ORCR, and the posterior aspect appears to be the site of maximum gapping. Specific attention should be directed toward maximizing initial fixation of larger rotator cuff tears, especially at the posterior aspect.

Details

Language :
English
ISSN :
1532-6500
Volume :
17
Issue :
5
Database :
MEDLINE
Journal :
Journal of shoulder and elbow surgery
Publication Type :
Academic Journal
Accession number :
18595743
Full Text :