Discrimination between Bacillus species by impedance analysis of individual dielectrophoretically positioned spores.

Authors :
Beck JD
Shang L
Li B
Marcus MS
Hamers RJ
Source :
Analytical chemistry [Anal Chem] 2008 May 15; Vol. 80 (10), pp. 3757-61. Date of Electronic Publication: 2008 Apr 11.
Publication Year :
2008

Abstract

We combine the use of dielectrophoretic positioning with electrical impedance measurements to detect and discriminate between individual bacterial spores on the basis of their electrical response. Using lithographically defined microelectrodes, we use dielectrophoresis to manipulate individual bacterial spores between the electrodes. The introduction of a single spore between the microelectrodes produces a significant change in electrical response that is species-dependent. When positioned between two electrodes and an AC voltage was applied, single spores caused current increases averaging 6.8 (+/-2.4) pA for Bacillus mycoides to 1.18 (+/-0.37) pA for Bacillus licheniformis. Using a mixture of spores of two different species, we demonstrate the ability to distinguish the species of individual spores in real time. This work demonstrates the feasibility of using impedance measurements for real-time detection and discrimination between different types of spores.

Details

Language :
English
ISSN :
1520-6882
Volume :
80
Issue :
10
Database :
MEDLINE
Journal :
Analytical chemistry
Publication Type :
Academic Journal
Accession number :
18407615
Full Text :
https://doi.org/10.1021/ac702113t