Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact.

Authors :
Daubert JP
Zareba W
Cannom DS
McNitt S
Rosero SZ
Wang P
Schuger C
Steinberg JS
Higgins SL
Wilber DJ
Klein H
Andrews ML
Hall WJ
Moss AJ
Source :
Journal of the American College of Cardiology [J Am Coll Cardiol] 2008 Apr 08; Vol. 51 (14), pp. 1357-65.
Publication Year :
2008

Abstract

Objectives: This study sought to identify the incidence and outcome related to inappropriate implantable cardioverter-defibrillator (ICD) shocks, that is, those for nonventricular arrhythmias.<br />Background: The MADIT (Multicenter Automatic Defibrillator Implantation Trial) II showed that prophylactic ICD implantation improves survival in post-myocardial infarction patients with reduced ejection fraction. Inappropriate ICD shocks are common adverse consequences that may impair quality of life.<br />Methods: Stored ICD electrograms from all shock episodes were adjudicated centrally. An inappropriate shock episode was defined as an episode during which 1 or more inappropriate shocks occurred; another inappropriate ICD episode occurring within 5 min was not counted. Programmed parameters for patients with and without inappropriate shocks were compared.<br />Results: One or more inappropriate shocks occurred in 83 (11.5%) of the 719 MADIT II ICD patients. Inappropriate shock episodes constituted 184 of the 590 total shock episodes (31.2%). Smoking, prior atrial fibrillation, diastolic hypertension, and antecedent appropriate shock predicted inappropriate shock occurrence. Atrial fibrillation was the most common trigger for inappropriate shock (44%), followed by supraventricular tachycardia (36%), and then abnormal sensing (20%). The stability detection algorithm was programmed less frequently in patients receiving inappropriate shocks (17% vs. 36%, p = 0.030), whereas other programming parameters did not differ significantly from those without inappropriate shocks. Importantly, patients with inappropriate shocks had a greater likelihood of all-cause mortality in follow-up (hazard ratio 2.29, p = 0.025).<br />Conclusions: Inappropriate ICD shocks occurred commonly in the MADIT II study, and were associated with increased risk of all-cause mortality.

Details

Language :
English
ISSN :
1558-3597
Volume :
51
Issue :
14
Database :
MEDLINE
Journal :
Journal of the American College of Cardiology
Publication Type :
Academic Journal
Accession number :
18387436
Full Text :
https://doi.org/10.1016/j.jacc.2007.09.073