Measuring noncalcified coronary atherosclerotic plaque using voxel analysis with MDCT angiography: phantom validation.
- Source :
-
AJR. American journal of roentgenology [AJR Am J Roentgenol] 2008 Apr; Vol. 190 (4), pp. W242-6. - Publication Year :
- 2008
-
Abstract
- Objective: This purpose of this study was to evaluate the accuracy and reproducibility of a voxel analysis technique for measuring noncalcified plaque in the coronary arteries.<br />Materials and Methods: Polyethylene phantoms representing noncalcified plaque were scanned in both MDCT and micro-CT scanners and inter- and intrareader variability of volume calculation was performed.<br />Results: Volume measurements by both MDCT and micro-CT were comparable to the true volume as measured by micrometry (< 3%, p = 0.05).<br />Conclusion: There appears to be no significant difference (< 3%) between MDCT and micro-CT measurements.