The predictive performance of a pharmacokinetic model for manually adjusted infusion of liquid sevofluorane for use with the Anesthetic-Conserving Device (AnaConDa): a clinical study.
- Source :
-
Anesthesia and analgesia [Anesth Analg] 2008 Apr; Vol. 106 (4), pp. 1207-14, table of contents. - Publication Year :
- 2008
-
Abstract
- Background: The Anesthetic-Conserving Device (AnaConDa) can be used to administer inhaled anesthetics using an intensive care unit (ICU) ventilator. We evaluated the predictive performance of a simple manually adjusted pump infusion scheme, for infusion of liquid sevoflurane to the AnaConDa.<br />Methods: We studied 50 ICU patients who received sevoflurane via the AnaConDa. They were randomly divided into three groups. A 6-h infusion of liquid anesthetic was adjusted according to the infusion scheme to a target end-tidal sevoflurane concentration of 1% (Group 1%, n = 15) and 1.5% (Group 1.5%, n = 15). The initial rate was adjusted to reach the target concentration in 10 min and then the infusion was reduced to the first hour maintenance rate and readjusted once each hour afterwards. The actual concentrations were measured in the breathing circuit and compared with the target values. In the third group (n = 20) we used the model to increase and decrease the target concentration (+/-0.3%) for 3 h and evaluated the actual change in concentration achieved. The ability of the infusion scheme to provide the target concentration was quantified by calculating the performance error (PE). Infusion scheme performance was evaluated in terms of accuracy (median absolute PE, MDAPE) and bias (median PE, MDPE).<br />Results: Performance parameters (mean +/- SD, %) were for 1%, 1.5%, increase of concentration by 0.3% and decrease of concentration by 0.3% groups, respectively: MDAPE 5.3 +/- 5.5, 2.6 +/- 4.0, 5.0 +/- 5.6, 5.5 +/- 5.4; MDPE -5.3 +/- 5.5, -2.3 +/- 4.1, -0.1 +/- 7.1, 0.2 +/- 5.4. No significant differences were found between means of all performance parameters when the 1% and 1.5% groups were compared.<br />Conclusions: There is an excellent 6-h predictive performance of a simplified pharmacokinetic model for manually adjusted infusion of liquid sevoflurane when using the AnaConDa to deliver sevoflurane to ICU patients.
- Subjects :
- Anesthetics, Inhalation administration & dosage
Blood Pressure
Computers
Humans
Infusions, Intravenous
Intensive Care Units
Kinetics
Metabolic Clearance Rate
Methyl Ethers administration & dosage
Models, Biological
Postoperative Period
Respiration, Artificial instrumentation
Respiration, Artificial methods
Sevoflurane
Tissue Distribution
Anesthetics, Inhalation pharmacokinetics
Methyl Ethers pharmacokinetics
Subjects
Details
- Language :
- English
- ISSN :
- 1526-7598
- Volume :
- 106
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Anesthesia and analgesia
- Publication Type :
- Academic Journal
- Accession number :
- 18349194
- Full Text :
- https://doi.org/10.1213/ane.0b013e31816782ff