Back to Search Start Over

The predictive performance of a pharmacokinetic model for manually adjusted infusion of liquid sevofluorane for use with the Anesthetic-Conserving Device (AnaConDa): a clinical study.

Authors :
Belda JF
Soro M
Badenes R
Meiser A
García ML
Aguilar G
Martí FJ
Source :
Anesthesia and analgesia [Anesth Analg] 2008 Apr; Vol. 106 (4), pp. 1207-14, table of contents.
Publication Year :
2008

Abstract

Background: The Anesthetic-Conserving Device (AnaConDa) can be used to administer inhaled anesthetics using an intensive care unit (ICU) ventilator. We evaluated the predictive performance of a simple manually adjusted pump infusion scheme, for infusion of liquid sevoflurane to the AnaConDa.<br />Methods: We studied 50 ICU patients who received sevoflurane via the AnaConDa. They were randomly divided into three groups. A 6-h infusion of liquid anesthetic was adjusted according to the infusion scheme to a target end-tidal sevoflurane concentration of 1% (Group 1%, n = 15) and 1.5% (Group 1.5%, n = 15). The initial rate was adjusted to reach the target concentration in 10 min and then the infusion was reduced to the first hour maintenance rate and readjusted once each hour afterwards. The actual concentrations were measured in the breathing circuit and compared with the target values. In the third group (n = 20) we used the model to increase and decrease the target concentration (+/-0.3%) for 3 h and evaluated the actual change in concentration achieved. The ability of the infusion scheme to provide the target concentration was quantified by calculating the performance error (PE). Infusion scheme performance was evaluated in terms of accuracy (median absolute PE, MDAPE) and bias (median PE, MDPE).<br />Results: Performance parameters (mean +/- SD, %) were for 1%, 1.5%, increase of concentration by 0.3% and decrease of concentration by 0.3% groups, respectively: MDAPE 5.3 +/- 5.5, 2.6 +/- 4.0, 5.0 +/- 5.6, 5.5 +/- 5.4; MDPE -5.3 +/- 5.5, -2.3 +/- 4.1, -0.1 +/- 7.1, 0.2 +/- 5.4. No significant differences were found between means of all performance parameters when the 1% and 1.5% groups were compared.<br />Conclusions: There is an excellent 6-h predictive performance of a simplified pharmacokinetic model for manually adjusted infusion of liquid sevoflurane when using the AnaConDa to deliver sevoflurane to ICU patients.

Details

Language :
English
ISSN :
1526-7598
Volume :
106
Issue :
4
Database :
MEDLINE
Journal :
Anesthesia and analgesia
Publication Type :
Academic Journal
Accession number :
18349194
Full Text :
https://doi.org/10.1213/ane.0b013e31816782ff