Back to Search
Start Over
Improving goniometer accuracy by compensating for individual transducer characteristics.
- Source :
-
Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology [J Electromyogr Kinesiol] 2009 Aug; Vol. 19 (4), pp. 704-9. Date of Electronic Publication: 2008 Mar 07. - Publication Year :
- 2009
-
Abstract
- Flexible goniometers are useful for direct movement measurements. Crosstalk due to rotation between the endblocks is well known. However, even without any rotation, some crosstalk can occur. The objective of this study was to elucidate the effect of, and compensate for, the inherent crosstalk in biaxial goniometers, with specific relevance for applications with one dominating movement direction. Six biaxial goniometers (M110, Biometrics Ltd., Gwent, UK) were evaluated. A precision jig, for simulating pure flexion/extension angles, was constructed. Each sensor produced a consistent and specific crosstalk pattern, when tested over a +/-100 degrees range of motion. A procedure for correction for the inherent crosstalk of individual goniometer, based on polynomial adjust, is presented. The method for compensation, which reduced the root mean square error from, on average for the six goniometers, 3.7 degrees (range 1.8-10.1 degrees) to 0.35 degrees (0.12-0.55 degrees), might be required for obtaining valid goniometer measurements, e.g. of valgus/varus of the knee during gait flexion/extension movements.
Details
- Language :
- English
- ISSN :
- 1873-5711
- Volume :
- 19
- Issue :
- 4
- Database :
- MEDLINE
- Journal :
- Journal of electromyography and kinesiology : official journal of the International Society of Electrophysiological Kinesiology
- Publication Type :
- Academic Journal
- Accession number :
- 18316206
- Full Text :
- https://doi.org/10.1016/j.jelekin.2008.01.006