In vivo measurements of ultrasound transmission through the human proximal femur.
- Source :
-
Ultrasound in medicine & biology [Ultrasound Med Biol] 2008 Jul; Vol. 34 (7), pp. 1186-90. Date of Electronic Publication: 2008 Feb 21. - Publication Year :
- 2008
-
Abstract
- Quantitative ultrasound (QUS) measurements can be used to estimate osteoporotic fracture risk. The commonly used variables are the speed of sound (SOS) and the frequency dependent sound attenuation (broadband ultrasound attenuation, [BUA]) of a wave propagating through the bone, preferably the calcaneus. The technology, so far, is less suitable for direct measurement in vivo at the spine or the femur for prediction of bone mineral density (BMD) or fracture risk at the main osteoporotic fracture sites. To improve the clinical performance of QUS, we built a device for direct QUS measurements at the human femur in vivo. In vivo images of ultrasound transmission at one of the main fracture sites, the proximal femur, could be acquired. The estimated precision of SOS measurements of 0.5% achieved at the femur is comparable with the precision of peripheral QUS devices.