Back to Search Start Over

[The measurement of radiation dosage in diagnostic x-ray device].

Authors :
Liu X
Source :
Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi [Sheng Wu Yi Xue Gong Cheng Xue Za Zhi] 2007 Dec; Vol. 24 (6), pp. 1256-9.
Publication Year :
2007

Abstract

In order to improve the diagnostic quality of X-ray device and to promote the implementation of radiation protection in accordance to the ICRU definition and ICRP classification standards for measuring radiation harm to human body, we have made a number of tests, including the kinetic energy released by ionizing air, the Half-value layer, the output dose coincidence, the High contrast resolution, the Consistency between visual light field and project X-ray field, and the Focus of X-ray tube. We have also analysed the measured data and results. The tests were conducted on the XEB150L-20 made by Shimadzu Co. in the middle of 1990s.

Details

Language :
Chinese
ISSN :
1001-5515
Volume :
24
Issue :
6
Database :
MEDLINE
Journal :
Sheng wu yi xue gong cheng xue za zhi = Journal of biomedical engineering = Shengwu yixue gongchengxue zazhi
Publication Type :
Academic Journal
Accession number :
18232472