Thin-section CT of lung without ECG gating: 64-detector row CT can markedly reduce cardiac motion artifact which can simulate lung lesions.
- Source :
-
European journal of radiology [Eur J Radiol] 2009 Jan; Vol. 69 (1), pp. 102-7. Date of Electronic Publication: 2007 Dec 19. - Publication Year :
- 2009
-
Abstract
- Purpose: Motion artifacts, which can mimic thickened bronchial wall and the cystic appearance of bronchiectasis, constitute a potential pitfall in the diagnosis of interstitial or bronchial disease. Therefore, purpose of our study was to evaluate whether 64-detector row CT (64-MDCT) enables a reduction in respiratory or cardiac motion artifacts in the lung area on thin-section CT without ECG gating, and to examine the correlation between cardiac motion artifact and heart rate.<br />Materials and Methods: Thirty-two patients with suspected diffuse lung disease, who underwent both 8- and 64-MDCT (gantry rotation time, 0.5 and 0.4s, respectively), were included. The heart rates of an additional 155 patients were measured (range, 48-126 beats per minute; mean, 76 beats per minute) immediately prior to 64-MDCT, and compared to the degree of cardiac motion artifact. Two independent observers evaluated the following artifacts on a monitor without the knowledge of relevant clinical information: (1) artifacts on 8- and 64-MDCT images with 1.25-mm thickness and those on 64-MDCT images with 0.625-mm thickness in 32 patients; and (2) artifacts on 64-MDCT images with 0.625-mm thickness in 155 patients.<br />Results: Interobserver agreement was good in evaluating artifacts on 8-MDCT images with 1.25-mm thickness (weighted Kappa test, kappa=0.61-0.71), and fair or poor in the other evaluations (kappa<0.31). Two observers stated that cardiac motion artifacts were more significant on 8-MDCT than on 64-MDCT in all 32 patients. Statistically significant differences were found at various checkpoints only in comparing artifacts between 8- and 64-MDCT for 1.25-mm thickness (Wilcoxon's signed-rank test, p<0.0017). Cardiac motion artifacts on 64-MDCT had no significant correlation with heart rate (Spearman's correlation coefficient by rank test).<br />Conclusion: The high temporal resolution of 64-MDCT appears to reduce cardiac motion artifact that can affect thin-section scans of the lung parenchyma.
- Subjects :
- Adult
Aged
Aged, 80 and over
Cardiac-Gated Imaging Techniques
Female
Humans
Image Enhancement instrumentation
Male
Middle Aged
Motion
Reproducibility of Results
Respiratory Mechanics
Retrospective Studies
Sensitivity and Specificity
Tomography, X-Ray Computed instrumentation
Artifacts
Image Enhancement methods
Lung Diseases diagnostic imaging
Tomography, X-Ray Computed methods
Subjects
Details
- Language :
- English
- ISSN :
- 1872-7727
- Volume :
- 69
- Issue :
- 1
- Database :
- MEDLINE
- Journal :
- European journal of radiology
- Publication Type :
- Academic Journal
- Accession number :
- 18068924
- Full Text :
- https://doi.org/10.1016/j.ejrad.2007.10.027