Oxygen isotope ratio measurements in CO(2) by means of a continuous-wave quantum cascade laser at 4.3 mum.
- Source :
-
Optics letters [Opt Lett] 2007 Oct 15; Vol. 32 (20), pp. 3047-9. - Publication Year :
- 2007
-
Abstract
- A mid-infrared laser spectrometer was developed for simultaneous high-precision (18)O/(16)O and (17)O/(16)O isotope ratio measurements in carbon dioxide. A continuous-wave, liquid-nitrogen cooled, distributed feedback quantum cascade laser, working at a wavelength of 4.3 microm, was used to probe (12)C(16)O(2), (16)O(12)C(18)O, and (16)O(12)C(17)O lines at ~2311.8 cm(-1). High sensitivity was achieved by means of wavelength modulation spectroscopy with second-harmonic detection. The experimental reproducibility in the short and long terms was deeply investigated through the accurate analysis of a large number of spectra. In particular, we found a short term precision of 0.5 per thousand and 0.6 per thousand, respectively, for (18)O/(16)O and (17)O/(16)O isotope ratios. The occurrence of systematic deviations is also discussed.
Details
- Language :
- English
- ISSN :
- 0146-9592
- Volume :
- 32
- Issue :
- 20
- Database :
- MEDLINE
- Journal :
- Optics letters
- Publication Type :
- Academic Journal
- Accession number :
- 17938695
- Full Text :
- https://doi.org/10.1364/ol.32.003047