Back to Search
Start Over
Analytical and clinical performance of the Ortho-Clinical Diagnostics VITROS amino-terminal pro-B type natriuretic peptide assay.
- Source :
-
Clinica chimica acta; international journal of clinical chemistry [Clin Chim Acta] 2008 Jan; Vol. 387 (1-2), pp. 48-54. Date of Electronic Publication: 2007 Sep 08. - Publication Year :
- 2008
-
Abstract
- Background: Measurement of amino-terminal pro-B-type natriuretic peptide (NT-proBNP) is useful for evaluating patients with heart failure (HF).<br />Methods: We evaluated the performance of a new automated NT-proBNP assay.<br />Results: The VITROS NT-proBNP assay had mean within-run and total imprecision of 1.0% and 3.4% at NT-proBNP concentrations from 67-27,500 ng/l. Acceptable linearity, functional/analytical sensitivity were demonstrated. Anticoagulant/tube types had no effect on results. Excellent sample stability and no high-dose hook were observed. High correlation between the VITROS and Elecsys methods was demonstrated (r=0.995; P<.001), with 98.3% clinical concordance. VITROS NT-proBNP concentrations were significantly higher in HF subjects than those without (1210 versus 68 ng/l; P<.001) and associated with HF symptom severity (P<.001). The VITROS assay had AUC for HF of 0.95 (P<.001), and had excellent NPV for excluding HF.<br />Conclusions: The automated VITROS NT-proBNP assay demonstrates excellent analytical and clinical performance for evaluating the presence and severity of HF.
Details
- Language :
- English
- ISSN :
- 0009-8981
- Volume :
- 387
- Issue :
- 1-2
- Database :
- MEDLINE
- Journal :
- Clinica chimica acta; international journal of clinical chemistry
- Publication Type :
- Academic Journal
- Accession number :
- 17915201
- Full Text :
- https://doi.org/10.1016/j.cca.2007.08.018