Direct stenting limits sirolimus-eluting stent edge neointimal thickening.
- Source :
-
Journal of vascular surgery [J Vasc Surg] 2007 Aug; Vol. 46 (2), pp. 354-9. - Publication Year :
- 2007
-
Abstract
- Objective: The use of sirolimus eluting stent (SES) has strongly limited the incidence of in-stent restenosis that still remains a problem at the stent edge. The aim of this study was to analyze the neointimal thickening after implantation of SES and to assess the influence of the stent implantation procedure on the neointimal thickening in the in-stent segment and at the edge of the stent in an ex-vivo model of stented human artery.<br />Methods: Both balloon expandable SES and the corresponding bare metal stent (BMS) were used in a model of human mammary artery culture. Stents were implanted either directly or after predilatation (10 atm, 60 seconds) and analysis of arterial segments were performed at 28 days poststenting. Cell proliferation and neointimal thickening were assessed by immunohistochemistry, western blotting, and histomorphometry, both in the in-stent segment and at the edge of the stent. Neointimal thickening was expressed as the ratio ([neointimal area/neointimal area + media area]).<br />Results: The in-stent neointimal thickening was dramatically inhibited in the SES group compared with the BMS group whatever the stenting technique was (predilatation: 0.22 +/- 0.05 vs 0.30 +/- 0.10; P < .04; direct stenting 0.16 +/- 0.04 vs 0.30 +/- 0.13; P <.01). This effect of SES was associated with a smallest expression of the small G protein RhoA and an increase of p27kip expression. In the BMS group, predilatation and direct stenting gave similar in-stent neointimal thickening. In contrast, in the SES group, in-stent neointimal thickening was significantly reduced when direct stenting was performed (0.16 +/- 0.04 [direct stenting] vs 0.22 +/- 0.05 [predilatation], P < .03). At the stent edge, a similar neointimal thickening was observed with both type of stent when predilatation was performed on the entire segment of the artery. Direct stenting significantly reduced the neointimal thickness at the stent edge when SES where used (0.06 +/- 0.01 [direct stenting] vs 0.19 +/- 0.06 [predilatation]; P < .001) but not in the BMS group.<br />Conclusion: These results confirm the efficiency of sirolimus released form SES to inhibit RhoA expression and to increase p27kip level in the arterial wall and show the benefit of direct stenting to limit the edge effect with SES.
- Subjects :
- Catheterization
Cyclin-Dependent Kinase Inhibitor p27 metabolism
Humans
Mammary Arteries metabolism
Mammary Arteries surgery
Mammary Arteries ultrastructure
Microscopy, Electron, Scanning
Organ Culture Techniques
Prosthesis Design
Time Factors
Tunica Intima metabolism
Tunica Intima surgery
Tunica Intima ultrastructure
rhoA GTP-Binding Protein metabolism
Blood Vessel Prosthesis Implantation instrumentation
Cardiovascular Agents administration & dosage
Cell Proliferation drug effects
Mammary Arteries drug effects
Metals
Sirolimus administration & dosage
Stents
Tunica Intima drug effects
Subjects
Details
- Language :
- English
- ISSN :
- 0741-5214
- Volume :
- 46
- Issue :
- 2
- Database :
- MEDLINE
- Journal :
- Journal of vascular surgery
- Publication Type :
- Academic Journal
- Accession number :
- 17664110
- Full Text :
- https://doi.org/10.1016/j.jvs.2007.04.063