Back to Search
Start Over
Mechanical evaluation of a new minimally invasive device for stabilization of proximal humeral fractures in elderly patients: a cadaver study.
- Source :
-
Acta orthopaedica [Acta Orthop] 2007 Jun; Vol. 78 (3), pp. 430-5. - Publication Year :
- 2007
-
Abstract
- Background: Treatment of proximal humerus fractures in elderly patients is challenging because of reduced bone quality. We determined the in vitro characteristics of a new implant developed to target the remaining bone stock, and compared it with an implant in clinical use.<br />Methods: Following osteotomy, left and right humeral pairs from cadavers were treated with either the Button-Fix or the Humerusblock fixation system. Implant stiffness was determined for three clinically relevant cases of load: axial compression, torsion, and varus bending. In addition, a cyclic varus-bending test was performed.<br />Results: We found higher stiffness values for the humeri treated with the ButtonFix system--with almost a doubling of the compression, torsion, and bending stiffness values. Under dynamic loading, the ButtonFix system had superior stiffness and less K-wire migration compared to the Humerusblock system.<br />Interpretation: When compared to the Humerusblock design, the ButtonFix system showed superior biomechanical properties, both static and dynamic. It offers a minimally invasive alternative for the treatment of proximal humerus fractures.
Details
- Language :
- English
- ISSN :
- 1745-3674
- Volume :
- 78
- Issue :
- 3
- Database :
- MEDLINE
- Journal :
- Acta orthopaedica
- Publication Type :
- Academic Journal
- Accession number :
- 17611860
- Full Text :
- https://doi.org/10.1080/17453670710014031