Back to Search Start Over

Will the hold of solid biodegradable implants be influenced by swelling during the degradation process? An in-vitro study with Meniscus Arrows.

Authors :
Wouters DB
Bos RR
De Hosson JT
Source :
Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA [Knee Surg Sports Traumatol Arthrosc] 2007 Oct; Vol. 15 (10), pp. 1204-9. Date of Electronic Publication: 2007 Jun 23.
Publication Year :
2007

Abstract

Water uptake after implantation of biodegradable devices induces swelling, as mentioned in literature. The hold in bone of solid devices will increase if the swelling is substantial enough. The results of weighing six Meniscus Arrows (MAs) before and after immersion in a sterile phosphate buffered saline solution during different time intervals were compared with the outcome of measurements under a field emission scanning electron microscope of six other MAs, stored under comparable conditions. The data were statistically evaluated with the Wilcoxon's signed rank test. The weight increase of 2.1 mg or 9.16% was statistically significant in the first 2 h following immersion, remaining stable afterwards with an average weight gain of 1.7 mg or 7.18%. The core diameter of the MAs increased to 0.01 mm or 1.01% with time. Although this is statistically significant, it is not expected to have any consequences for the hold. However, a remarkable and statistically significant decrease in the outer inter-barb diameter of 0.15 mm or 8.6% was noted with time. Mechanical testing should reveal the clinical relevance of the results of this study.

Details

Language :
English
ISSN :
0942-2056
Volume :
15
Issue :
10
Database :
MEDLINE
Journal :
Knee surgery, sports traumatology, arthroscopy : official journal of the ESSKA
Publication Type :
Academic Journal
Accession number :
17589827
Full Text :
https://doi.org/10.1007/s00167-007-0366-0